Evaluation of the Carbon Isotopic Effects of NDIR and CRDS Analyzers on Atmospheric CO 2 Measurements

LingJun Xia,LingXi Zhou,Marcel V. van der Schoot,Chris W. Rella,LiXin Liu,Gen Zhang,HongYang Wang
DOI: https://doi.org/10.1007/s11430-016-5294-8
2016-01-01
Abstract:Non-dispersive infrared (NDIR) and cavity ring-down spectroscopy (CRDS) CO 2 analyzers use 12 CO 2 isotopologue absorption lines and are insensitive to all or part of other CO 2 -related isotopologues. This may produce biases in CO 2 mole fraction measurements of a sample if its carbon isotopic composition deviates from that of the standard gases being used. To evaluate and compare the effects of carbon isotopic composition on NDIR and CRDS CO 2 analyzers, we prepared three test sample air cylinders with varying carbon isotopic abundances and calibrated them against five standard cylinders with ambient carbon isotopic composition using CRDS and NDIR systems. We found that the CO 2 mole fractions of the sample cylinders measured by G1301 (CRDS) were in good agreement with those measured by LoFlo (NDIR). The CO 2 values measured by both instruments were higher than that of a CO 2 isotope measured by G2201i (CRDS) analyzer for a test cylinder with depleted carbon isotopic composition δ 13 C =-36.828‰, whereas no obvious difference was found for other two test cylinders with δ 13 C=-8.630‰ and δ 13 C=-15.380‰, respectively. According to the theoretical and experimental results, we concluded that the total CO 2 mole fractions of samples with depleted isotopic compositions can be corrected on the basis of their 12 CO 2 values calibrated by standard gases using LoFlo and G1301 if the δ 13 C and δ 18 O values are known.