Determination of 7 Indictor Polychlorinated Biphenyls (pcbs) Residues in Porphyra by Ultrasonic Extraction and Gas Chromatography (GC)

Liangliang Tian,Feng Han,Youqiong Cai,Cong Kong,Yongfu Shi,Yuan Wang,Guangxin Yang,Qianyun Zhan,Dongmei Huang
DOI: https://doi.org/10.4236/jacen.2016.51b001
2016-01-01
Journal of Agricultural Chemistry and Environment
Abstract:A method was developed for determination of 7 indictors the polychlorinated biphenyls (PCBs) residues in porphyra by gas chromatography (GC) with electron capture detector (ECD). The PCBs were extracted with hexane/methylene chloride (1:1, v/v) by ultrasonic extraction and the samples were cleaned up by concentrated sulfuric acid and Alumina-N solid phase extraction cartridge. The analytes were quantified by an internal standard method. Under optimal experimental conditions, good linearity was observed in the range of 5 - 200 ng/mL and the correlation coefficients were 0.9994 - 0.9998. The limit of quantitation (LOQ) for target analytes ranged from 6.0 to 10.0 μg/kg. At the spiked levels of 10, 50, 100 μg/kg, the average recoveries ranged from 90.9% to 102% with the relative standard deviations 2.12% - 6.32%. The result showed that the proposed method was rapid, and could be used for the determination of the PCBs in porphyra.