Long-term Results of Total Hip Arthroplasty with Chinese Porous Coated Anatomic Prothesis: a 11-to 17-Year Follow-Up Study

林剑浩,吕厚山,寇伯龙,袁燕林,钟群杰
DOI: https://doi.org/10.3760/cma.j.issn.0253-2352.2010.02.007
2010-01-01
Abstract:Objective To evaluate the long-term results of total hip arthroplasty with Chinese porous coated anatomic prosthesis (11-17 years). Methods Ninty-five patients who received total hip arthroplasty with Chinese porous coated anatomic prosthesis in our center from January 1990 to April 1998 were followed up retrospectively. The clinical results analyzed with Harris hip score, and the roentgenographic results ana-lyzed with DeLee and Charnley methods, respectively. Results Of the 95 patients, 50 patients lost contact,and 16 patients died. The rest (14 males and 15 females, 37 hips) were reviewed with an average of 13.8 years follow-up. The mean Harris hip score was 70.3 (6-98). 48.5% of them were rated as excellent and good (16 hips), 30.3% of them rated as bad. 6 hips of them needed revision, mainly due to aseptic loosening. 18 cases of them considered that the pain disappeared or slight pain that didn't affect the function of the hip joints exists after the primary replacement. The roentgenogram showed that for the femoral components, 9 femoral stems were subsided by 9.3 mm on average, osteolysis occurred in 10 hips, radiolucency around the stem occurred in 12 hips, and for the acetabular components, radiolucency occurred in 13 hips, the porous coated layers shed in 2 hips. Conclusion The long-term results of total hip arthroplasty with Chinese porous coated anatomic prothesis done in our center during 1990 to 1999 are fair, and aseptic loosening is the main complication.
What problem does this paper attempt to address?