Cable Shared Dual-Frequency Catheter for Intravascular Ultrasound.

Min Su,Zhiqiang Zhang,Jiehan Hong,Yaocai Huang,Peitian Mu,Yanyan Yu,Rong Liu,Suzi Liang,Hairong Zheng,Weibao Qiu
DOI: https://doi.org/10.1109/tuffc.2019.2898256
IF: 3.267
2019-01-01
IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
Abstract:This study proposes a catheter consisting of dual-frequency transducer for intravascular ultrasound. Both ultrasonic elements with different frequencies were connected to one coaxial cable to make the connection simple. The aperture sizes of the ultrasound elements were 0.4 x 0.6 and 0.3 x 0.4 mm 2 for the low-frequency element and high-frequency element, respectively. The center frequency and bandwidth of the fabricated low-frequency transducer were 33.8 MHz and 49.3%, respectively. Meanwhile, the center frequency and bandwidth of the high-frequency transducer were 80.6 MHz and 50.3%, respectively. Imaging evaluations of wire phantom, tissue phantom, and vessel tissue demonstrated good imaging capability of the dual-frequency catheter. The spatial resolutions are 19 mu m axially and 128 mu m laterally for the high-frequency transducer, and 37 mu m axially and 199 mu m laterally for the low-frequency transducer. Bandpass filters were designed to separate the mixed echo signals. After filtering, the images from different ultrasound elements can be successfully identified, indicating the feasibility of the proposed cable-shared dual-frequency imaging strategy. The proposed method has simple structure, good imaging resolution, and large penetration depth, showing good application prospect for intravascular ultrasound.
What problem does this paper attempt to address?