A Quantitative Comparison Of Five Optical Coherence Tomography Angiography Systems In Clinical Performance

Xin-Xin Li,Wei Wu,Hao Zhou,Jun-Jie Deng,Meng-Ya Zhao,Tian-Wei Qian,Chen Yan,Xun Xu,Su-Qin Yu
DOI: https://doi.org/10.18240/ijo.2018.11.09
IF: 1.645
2018-01-01
International Journal of Ophthalmology
Abstract:AIM: To compare the clinical performance of 4 spectraldomain (SD) optical coherence tomography angiography (OCTA) systems: AngioVue (TM), AngioPlex (TM), Spectralis (R) OCTA, AngioScan, and 1 swept-source (SS) OCTA SS OCT Angio (TM).METHODS: Twenty-seven undilated right eyes of 27 participants underwent OCTA examination using five different systems respectively for both 3x3 and 6x6 mm(2) scan pattern (Spectralis OCTA for 3x3 mm' scan only). Image quality, including vessel valid visibility and the number of motion artifacts, and acquisition time were evaluated. Repeated measures analysis of variance (ANOVA) with Bonferroni's post-test and Friedman test with Dunn's post-test were used to compare measurements.RESULTS: The age of the subjects was 28.19 +/- 5.55y (range, 23-49y). The spherical equivalent refraction was -2.55 +/- 1.84 D (range, 0.00 to -5.25 D). Significant difference was observed in the evaluation of vessel valid visibility (AngioVue the highest: 0.111 +/- 0.031 for 3x3 mm(2) scan and 0.128 +/- 0.020 for 6x6 mm(2) scan), number of motion artifacts (AngioVue the fewest: 0.778 +/- 1.086 for 3x3 mm(2) scan and 0.333 +/- 0.620 for 6x6 mm(2) scan) and acquisition time (AngioPlex the shortest: 8.537 +/- 1.921s for 3x3 mm(2) scan and 8.298 +/- 1.741s for 6x6 mm(2) scan; all P<0.001).CONCLUSION: There is poor agreement of measurements among systems. AngioVue provides images with the highest vessel valid visibility and the fewest motion artifacts. AngioPlex achieves the shortest acquisition.
What problem does this paper attempt to address?