OCT-angiography: A qualitative and quantitative comparison of 4 OCT-A devices

Marion R. Munk,Helena Giannakaki-Zimmermann,Lieselotte Berger,Wolfgang Huf,Andreas Ebneter,Sebastian Wolf,Martin S. Zinkernagel
DOI: https://doi.org/10.1371/journal.pone.0177059
IF: 3.7
2017-05-10
PLoS ONE
Abstract:PURPOSE: To compare the quality of four OCT-angiography(OCT-A) modules.METHOD: The retina of nineteen healthy volunteers were scanned with four OCT-devices (Topcon DRI-OCT Triton Swept-source OCT, Optovue RTVue-XR, a prototype Spectralis OCT2, Heidelberg-Engineering and Zeiss Cirrus 5000-HD-OCT). The device-software generated en-face OCT-A images of the superficial (SCP) and deep capillary plexuses (DCP) were evaluated and scored by 3 independent retinal imaging experts. The SCP vessel density was assessed using Angiotool-software. After the inter-grader reliability assessment, a consensus grading was performed and the modules were ranked based on their scoring.RESULTS: There was no significant difference in the vessel density among the modules (Zeiss 48.7±4%, Optovue 47.9±3%, Topcon 48.3±2%, Heidelberg 46.5±4%, p = 0.2). The numbers of discernible vessel-bifurcations differed significantly on each module (Zeiss 2±0.9 bifurcations, Optovue 2.5±1.2, Topcon 1.3±0.7 and Heidelberg 0.5±0.6, p≤0.001). The ranking of each module differed depending on the evaluated parameter. In the overall ranking, the Zeiss module was superior and in 90% better than the median (Bonferroni corrected p-value = 0.04). Optovue was better than the median in 60%, Topcon in 40% and Heidelberg module in 10%, however these differences were not statistically significant.CONCLUSION: Each of the four evaluated OCT-A modules had particular strengths, which differentiated it from their competitors.
multidisciplinary sciences
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to conduct a comparative analysis of the quality of four different optical coherence tomography angiography (OCT - A) modules. Specifically, the research objective is to evaluate and compare the performance of four commercially available OCT - A modules in healthy subjects. These modules are Topcon DRI - OCT Triton Swept - source OCT, Optovue RTVue - XR, Spectralis OCT2 prototype (Heidelberg Engineering), and Zeiss Cirrus 5000 - HD - OCT. Through quantitative and qualitative methods, the study evaluated the image quality of the superficial capillary plexus (SCP) and the deep capillary plexus (DCP), aiming to determine the advantages and disadvantages of different devices in imaging quality, so as to provide references for clinical applications.