Correlation Polarimetry For Broadband Magnetic Fluctuation Measurements

E Parke,W X Ding,D L Brower
DOI: https://doi.org/10.1063/1.5038832
IF: 1.6
2018-01-01
Review of Scientific Instruments
Abstract:Correlation techniques have been successfully utilized for plasma diagnostics like electron cyclotron emission to reduce measurement noise. We present the first application of such a technique to Faraday-effect polarimetry measurements on the Madison Symmetric Torus (MST). TheMSTfar infrared (FIR) interferometer-polarimeter diagnostic utilizes 11 vertical chords with a chord separation of 7-8 cm and a heterodyne detection system for fluctuation measurements up to several hundred kHz. The planar-diode mixers viewing each chord represent independent noise sources; modifying the optical setup so that two different mixers view the same chord allows cross correlation between the two independent signals to reduce the noise floor in fluctuation measurements. In this manner, the noise floor in both interferometry and polarimetry measurements in reversed-field pinch discharges has been reduced by a factor of 20-30. The correlation polarimeter provides a sensitive measurement of broadband magnetic fluctuations. Published by AIP Publishing.