Endovascular Repair of an Aortic Arch Pseudoaneurysm by an Atrial Septal Defect Occluder Combined with a Chimney Stent
Guangqi Chang,Wei Chen,Hao Yin,Zilun Li,Xiaoxi Li,Shenming Wang
DOI: https://doi.org/10.1016/j.jvs.2012.10.095
IF: 4.86
2013-01-01
Journal of Vascular Surgery
Abstract:An aortic arch pseudoaneurysm distal to the left common carotid artery (LCCA) was repaired endovascularly with an atrial septal defect occluder. Aortography showed complete closure of the pseudoaneurysm but delayed blood flow in the LCCA. A chimney stent was then placed into the LCCA to salvage it. Completion angiography suggested free flow in the LCCA and a minor perioccluder leak, which disappeared at 1-month follow-up. The patient was free from any symptoms related to the pseudoaneurysm at 1-year follow-up, with the atrial septal defect occluder in a satisfactory position, absence of leaking, and complete thrombosis in the pseudoaneurysm cavity. An aortic arch pseudoaneurysm distal to the left common carotid artery (LCCA) was repaired endovascularly with an atrial septal defect occluder. Aortography showed complete closure of the pseudoaneurysm but delayed blood flow in the LCCA. A chimney stent was then placed into the LCCA to salvage it. Completion angiography suggested free flow in the LCCA and a minor perioccluder leak, which disappeared at 1-month follow-up. The patient was free from any symptoms related to the pseudoaneurysm at 1-year follow-up, with the atrial septal defect occluder in a satisfactory position, absence of leaking, and complete thrombosis in the pseudoaneurysm cavity. Thoracic pseudoaneurysms represent a rare variety of aortic disorders, with patients often asymptomatic, are potentially fatal, and are traditionally treated by open surgery. However, surgical procedure is often complicated, with poor outcomes and high morbidity and mortality.1Mulder E.J. Bockel H. Maas J. Morbidity and mortality of reconstructive surgery of noninfected false aneurysms detected long after aortic prosthetic reconstruction.Arch Surg. 1998; 133: 45-49Crossref PubMed Scopus (122) Google Scholar For the past 2 decades, endovascular therapy has been emerging as an effective and less invasive treatment for aortic disorders. Although conventional stent grafts have been widely used in aortic segments with adequate landing zones, branched grafts, fenestrated grafts, chimney grafts, and occluders have also been applied in challenging zones such as the aortic arch. In 2005, Bashir et al2Bashir F. Quaife R. Carroll J.D. Percutaneous closure of ascending aortic pseudoaneurysm using Amplatzer septal occluder device: the first clinical case report and literature review.Catheter Cardiovasc Interv. 2005; 65: 547-551Crossref PubMed Scopus (42) Google Scholar reported the first case of using a septal occluder device to close an ascending aortic pseudoaneurysm. In 2009, Hussain et al3Hussain J. Strumpf R. Wheatley G. Diethrich E. Percutaneous closure of aortic pseudoaneurysm by Amplatzer occluder device-case series of six patients.Catheter Cardiovasc Interv. 2009; 73: 521-529Crossref PubMed Scopus (68) Google Scholar reported six cases, so far the largest series, of thoracic pseudoaneurysm treated by endovascular placement of occluders. In 2009, Schmauss et al4Schmauss D. Sodian R. Schmitz C. Bigdeli A. Schmoeckel M. Lueth T. et al.3D-printing models to create custom-made devices for coil embolization of anastomotic leak after aortic arch replacement.Thorac Cardiovasc Surg. 2009; 88: 974-978Google Scholar reported the first case of aortic arch pseudoaneurysm successfully treated by an occluder. In this report, we describe a case of aortic arch pseudoaneurysm that was successfully treated by an atrial septal defect (ASD) occluder combined with a chimney stent. To our knowledge, this is the first case published. The patient was a 41-year-old male with a chief complaint of chest and back pain, hoarseness, and dysphagia after blunt trauma. An aortic arch pseudoaneurysm (neck, 18 mm; size, 43 mm × 40 mm) right distal and posterior to the orifice of the left common carotid artery (LCCA) was revealed by computed tomography angiography (CTA) (Fig 1, A-C). Previously, we have shown at our facility that sealing the primary tear of a type B aortic dissection with a ventricular septal defect occluder has been successfully achieved.5Chang G.Q. Wang H.S. Chen W. Yao C. Li Z.L. Wang S.M. Endovascular repair of a type b aortic dissection with a ventricular septal defect occluder.J Vasc Surg. 2010; 51: 1507-1509Abstract Full Text Full Text PDF PubMed Scopus (10) Google Scholar We then considered treating this aortic arch pseudoaneurysm with an ASD occluder. As the disc of the occluder might cover the LCCA considering its close proximity to the neck of the pseudoaneurysm, the chimney technique was planned to salvage the LCCA in case unintentional blockage occurred. Benefits and risks of the procedure were fully explained to the patient and family members before the informed consent was signed. The procedure was performed under general anesthesia. Right femoral access was obtained with a 6F introducer sheath, through which 5000 IU of heparin was injected, followed by advancing a 0.035-inch hydrophilic guidewire and 5F pigtail catheter into the ascending aorta. A 45° left anterior oblique aortogram revealed an aortic arch pseudoaneurysm (neck, 18 mm; size, 43 × 40 mm) right distal to the orifice of the LCCA (Fig 1, D). The pigtail catheter was exchanged for a 5F Cobra catheter (Cordis Corporation, Bridgewater, NJ), which was then advanced through the neck into the pseudoaneurysm cavity. The 5F Cobra catheter was exchanged for a 0.035-inch Amplaze guidewire (Cordis Corporation), through which a 14F sheath was navigated into the cavity. A SearCareTM ASD occluder (38-24-34 mm; Lifetech Scientific Co, Ltd, Shenzhen, China) was advanced into the sheath and then the cavity. The anterior disc of the occluder was deployed in the cavity and gently pulled back against the rim of the cavity. The delivery catheter was further retracted to release the middle of the device across the neck and then the posterior disc positioned in the aorta. A subsequent aortogram showed complete closure of the pseudoaneurysm but delayed blood flow in the LCCA (Fig 2, A), which suggested obstruction of the LCCA. According to the preoperative plan, the chimney technique was applied to salvage the LCCA. LCCA access was obtained through cutdown. A 6F introducer sheath, 5F Cobra catheter, and a 0.035-inch hydrophilic guidewire were advanced into the LCCA retrogradely in sequence. An 8- × 60-mm self-expanding Zilver stent (Cook Medical, Bloomington, Ind) was then placed into the LCCA, with its proximal end 6 mm within the aortic arch. Completion angiography showed free flow in the LCCA and correct positioning of the occluder with minor leak (Fig 2, B), which was expected to disappear once the occluder was endothelialized and the cavity was thrombosed. The patient underwent an uneventful recovery without any transient or permanent neurologic complication postoperatively. CTA at 1 week (Fig 3, A and B) showed proper positioning of the occluder, although a minor leak remained. CTA at 3 months and 1 year (Fig 3, C-E) demonstrated complete occlusion of the pseudoaneurysm and patent graft in the LCCA. The patient was placed on long-term intake of aspirin (100 mg/d) and was free from any symptoms related to the pseudoaneurysm at 1-year follow-up. Thoracic pseudoaneurysms are conventionally managed by open surgery, which is associated with high morbidity and mortality.1Mulder E.J. Bockel H. Maas J. Morbidity and mortality of reconstructive surgery of noninfected false aneurysms detected long after aortic prosthetic reconstruction.Arch Surg. 1998; 133: 45-49Crossref PubMed Scopus (122) Google Scholar In the past 2 decades, endovascular management has increasingly become a less invasive and effective treatment for aortic disorders including thoracic pseudoaneurysm. However, for pseudoaneurysms in the aortic arch, conventional endovascular stenting can be difficult. Although branched stent grafts have been used in this challenging zone,6Chuter T.A. Schneider D.B. Endovascular repair of the aortic arch.Perspect Vasc Surg Endovasc Ther. 2007; 19: 188-192Crossref PubMed Scopus (45) Google Scholar they are time-consuming and expensive. Furthermore, the additional complexity of branched graft deployment may result in high rates of stroke. 6Chuter T.A. Schneider D.B. Endovascular repair of the aortic arch.Perspect Vasc Surg Endovasc Ther. 2007; 19: 188-192Crossref PubMed Scopus (45) Google Scholar Instead, endovascular placement of occluders has been emerging as a promising strategy to treat the pseudoaneurysm in proximity to supra-aortic vessels.2Bashir F. Quaife R. Carroll J.D. Percutaneous closure of ascending aortic pseudoaneurysm using Amplatzer septal occluder device: the first clinical case report and literature review.Catheter Cardiovasc Interv. 2005; 65: 547-551Crossref PubMed Scopus (42) Google Scholar, 3Hussain J. Strumpf R. Wheatley G. Diethrich E. Percutaneous closure of aortic pseudoaneurysm by Amplatzer occluder device-case series of six patients.Catheter Cardiovasc Interv. 2009; 73: 521-529Crossref PubMed Scopus (68) Google Scholar, 4Schmauss D. Sodian R. Schmitz C. Bigdeli A. Schmoeckel M. Lueth T. et al.3D-printing models to create custom-made devices for coil embolization of anastomotic leak after aortic arch replacement.Thorac Cardiovasc Surg. 2009; 88: 974-978Google Scholar Unlike tubular stent grafts, which require sufficient proximal and distal landing zones, proximity to supra-aortic vessels does not seem to be an issue for occluders. In the present case, the aortic arch pseudoaneurysm with a moderate neck was obliterated by an ASD occluder. However, the LCCA was also covered because of the extremely narrow space between the LCCA and the pseudoaneurysm neck. According to the preoperative plan, a chimney stent was deployed to salvage the LCCA. The chimney technique has been used to rescue vital branch vessels that are covered unintentionally.7Ohrlander T. Sonesson B. Ivancev K. Resch T. Dias N. Malina M. The chimney graft: a technique for preserving or rescuing aortic branch vessels in stent graft sealing zones.J Endovasc Ther. 2008; 15: 427-432Crossref PubMed Scopus (285) Google Scholar Although the occluder and chimney stent have been applied separately in aortic arch procedures, their combination as a novel treatment has not been reported previously. As both instruments are commercially available, they can be integrated to treat pseudoaneurysm in proximity to supra-aortic branch vessels. Although this complete endovascular strategy avoids significant injuries associated with surgery, it raises several concerns. In the largest series of thoracic pseudoaneurysm treated by endovascular placement of occluders, three minor leaks occurred among six patients. However, two of them resolved spontaneously.3Hussain J. Strumpf R. Wheatley G. Diethrich E. Percutaneous closure of aortic pseudoaneurysm by Amplatzer occluder device-case series of six patients.Catheter Cardiovasc Interv. 2009; 73: 521-529Crossref PubMed Scopus (68) Google Scholar Thus, most of the perioccluder leaks may disappear once the occluder endothelializes and the cavity thromboses. In the present case, the usage of a chimney stent may also increase the risk of leak. Notably, the chimney technique we performed is not a standard one that is composed of a large aortic stent graft and a smaller parallel chimney graft, which becomes tightly embraced by the larger aortic stent graft, thus, reducing the incidence of proximal type I endoleak. Although persistent leak is absent in our case, it remains unknown if the shorter “gutters” between the occluder and the chimney stent will increase the risk of leakage. Another concern is possible maldeployment of the occluder.3Hussain J. Strumpf R. Wheatley G. Diethrich E. Percutaneous closure of aortic pseudoaneurysm by Amplatzer occluder device-case series of six patients.Catheter Cardiovasc Interv. 2009; 73: 521-529Crossref PubMed Scopus (68) Google Scholar Furthermore, recurrence of an aortic pseudoaneurysm may occur.8Hussain J. Strumpf R. Ghandforoush A. Wheatley G. Sutherland J. Transcatheter closure of recurrent aortic pseudoaneurysm previously treated by Amplatzer occluder device.J Vasc Surg. 2010; 52: 196-198Abstract Full Text Full Text PDF PubMed Scopus (13) Google Scholar Thus, although the combination of the occluder and chimney stent may extend and simplify the application of endovascular techniques for treating aortic arch pseudoaneurysms, concerns about potential complications and durability have to be addressed before broader application can be advocated. In summary, management of aortic arch pseudoaneurysm by an ASD occluder, with secondary chimney stenting for rescuing the potentially unintentional coverage of branch vessels, seems technically feasible, safe, and effective at short-term follow-up. However, longer follow-up is needed to assess the efficacy and durability.