Study On A Step-Advanced Filter Monitor For Continuous Radon Progeny Measurement

Lei Zhang,Jinmin Yang,Qiuju Guo
DOI: https://doi.org/10.1093/rpd/ncw333
2017-01-01
Radiation Protection Dosimetry
Abstract:Traditional fixed-filter radon progeny monitors are usually clogged with the loading of dust and cannot be used for radon progeny continuous measurement for long period. To solve this problem, a step-advanced filter (SAF) monitor for radon progeny measurement was developed. This monitor automatically roll and stop the filter at each interview. Radon progeny is collected on a 'fresh' filter at a flowrate of 3 L/min. At the same time, alpha and beta particles emitted from filter are recorded by a PIPS detector. A newly developed alpha-beta spectrum method was used for radon progeny concentration calculation. The Po-218, Pb-214 and Bi-214 concentrations as well as equilibrium equivalent concentration (EEC) could be worked out at the same time. The lower level limit detection of this monitor is 0.48 Bq m(-3) (EEC) for 1h interval. Comparison experiments were carried out in the radon chamber at the National Institute of Metrology of China. The measurement results of this SAF monitor are consistent with EQF3220 (SARAD GmbH, Germany), and the uncertainty is smaller. Due to its high sensitivity, the periodical variation of radon progeny concentration can be easily observed by this monitor. The SAF moniter is suitable for continuous measurement in both indoor and outdoor environments.
What problem does this paper attempt to address?