Production of Medically Useful Nitric Monoxide Using AC Arc Discharge

S. R. Li,Y. F. Huang,Z. Liu,M. H. Sui,J. M. Liu,K. P. Yan
DOI: https://doi.org/10.1016/j.niox.2017.06.006
IF: 4.898
2018-01-01
Nitric Oxide
Abstract:Inhaled nitric monoxide (iNO) is increasingly used as a medical treatment for acute respiratory distress syndrome. A course of the existing nitric monoxide (NO) therapy with gas cylinders could cost up to approximately $15,000 for an average of 30.2 h. Moreover, a gas cylinder containing a mixture of N-2 and NO may potentially leak NO. The objective of this study is to develop an efficient and cost-effective on site iNO generation system. In the present setup, NO was generated by using dry air or mixed oxygen/nitrogen (O-2/N-2) and an AC power source with an output power level of 5-30 W at atmospheric pressure. The simultaneously produced NO2 was eliminated with an ammonium sulfite ((NH4)(2)SO3) solution. The effects of the O-2/N-2 ratio, gas flow rate, discharge gap distance, output energy density and electrode structure on NOx concentration and the NO/NO2 ratio are reported. The concentrations of NO and NO2 reached 62 ppm and 3 ppm, respectively, after absorption and dilution at a gas flow rate of 6 L/min. With the present setup, the AC arc discharge produced NOx at a stable concentration for at least 6 h using dry air. (C) 2017 Elsevier Inc. All rights reserved.