Research on Real-Time Self-Adjusting Calibration Algorithm for Glucose Sensor

Huang Yonghong,Liu Hongying,Tian Senfu,Pi Xitian
DOI: https://doi.org/10.3969/j.issn.0254-3087.2016.09.017
2016-01-01
Abstract:Aiming at the problem that subcutaneously implanted glucose sensor is susceptible to physiological environment influence in actual dynamic monitoring process,which restricts the measurement accuracy of the sensor,a real-time self-adjusting calibration algorithm based on one-point calibration technique is proposed in this paper,the signal drift compensation and sensitivity attenuation calibration of the sensor are achieved.This method can adjust the key parameters (including sensitivity and current data)in one-point calibration process,and sequentially establishes the optimal model.The least squares method is used to fit the initial parameters of the model.The calibration for all current data monitored in vivo for three days was exactly implemented based on the optimal model. Calibration results show that the relative error of the glucose concentration estimated using the method is within the range of 0.01 % to 6.67%,which is lower than the scope of 6.36% to 1 6.92% using the traditional one-point calibration model.The result indicates that the calibration accuracy and practical value are both improved.The proposed real-time self-adjusting calibration algorithm effectively improves the accuracy of continuous glucose monitoring.