Calibration algorithms for continuous glucose monitoring systems based on interstitial fluid sensing

Tianyi Sun,Jentsai Liu,Ching Jung Chen
DOI: https://doi.org/10.1016/j.bios.2024.116450
IF: 12.545
2024-05-30
Biosensors and Bioelectronics
Abstract:Continuous glucose monitoring (CGM) is of great importance to the treatment and prevention of diabetes. As a proven commercial technology, electrochemical glucose sensor based on interstitial fluid (ISF) sensing has high sensitivity and wide detection range. Therefore, it has good promotion prospects in noninvasive or minimally-invasive CGM system. However, since there are concentration differences and time lag between glucose in plasma and ISF, the accuracy of this type of sensors are still limited. Typical calibration algorithms rely on simple linear regression which do not account for the variability of the sensitivity of sensors. To enhance the accuracy and stability of CGM based on ISF, optimization of calibration algorithm for sensors is indispensable. While there have been considerable researches on improving calibration algorithms for CGM, they have still received less attention. This article reviews the problem of typical calibration and presents the outstanding calibration algorithms in recent years. Finally, combined with existing research and emerging sensing technologies, this paper makes an outlook on the future calibration algorithms for CGM sensors.
biotechnology & applied microbiology,biophysics,chemistry, analytical,electrochemistry,nanoscience & nanotechnology
What problem does this paper attempt to address?