The Comparison Of Electric Fields Distribution Applying Various Coil Configurations In Deep Transcranial Magnetic Stimulation

Xile Wei,Yao Li,Guosheng Yi,Meili Lu,Jiang Wang,Zhen Zhang
DOI: https://doi.org/10.1109/cisp-bmei.2017.8302255
2017-01-01
Abstract:Deep Transcranial magnetic stimulation (dTMS) is a promising noninvasive treatment of several neurological and psychiatric disorders. However, the distribution properties of electric fields induced by different stimulating coils are still unknown. In this paper, Crown-coils, Figure-8 coils and multi-coil arrays of various parameter settings were numerically used. The attenuation rate of induced electric field and activating focality in the spherical head model were calculated by finite element method. Simulation results show that the size and spatial arrangement of coil configurations have relatively great effects on induced-field attenuation rate and focality. Compared to Crown-coils and Figure-8 coils, multi-coil arrays tend to penetrate deeper brain regions at the expense of stimulating focality. It is also indicated that larger coils have slower field attenuation rate than that of smaller coils with similar topological structures. This work verified the trade-off between field attenuation rate and stimulation focality, indicating that coil settings for a specific application should evaluate both stimulation depth and focality.