A COD measurement method with turbidity compensation based on variable radial basis function neural network

Renhao Fan,Senlin Wang,Hao Chen
DOI: https://doi.org/10.1039/d3ay01537h
IF: 3.1
2023-09-28
Analytical Methods
Abstract:In recent years, ultraviolet-visible spectrometry has been widely used to measure sewage's Chemical oxygen demand (COD). However, most methods that use UV-vis spectroscopy for COD measurement have not eliminated the influence of turbidity. Therefore, this article proposes a new COD measurement method using UV-vis spectroscopy. This method includes a new turbidity compensation algorithm and an algorithm for COD measurement using the variable radial basis function (VRBF) neural network. Our turbidity compensation algorithm first utilizes principal component analysis (PCA) to extract the characteristic wavelengths of the spectrum. Then, turbidity is used to fit the absorbance difference caused by turbidity in the characteristic wavelength, and a turbidity compensation model is obtained. Use the turbidity compensation model to compensate for the absorption spectrum to remove the influence of turbidity on COD measurement. Secondly, the VRBF neural network model is used to measure COD concentration. The results show that, compared with the traditional Partial least squares regression model, the R2 coefficient of determination increases from 0.27 to 0.88, and the root mean square deviation decreases from 5.56 to 1.69. Compared with improved bagging algorithm and MLP algorithm, this method can measure COD concentration from absorption spectra faster, more directly, and more accurately.
spectroscopy,chemistry, analytical,food science & technology
What problem does this paper attempt to address?