A Route to Large-Scale Ultra-Low Noise Detector Arrays for Far-Infrared Space Applications

D J Goldie,S. Withington,C. N. Thomas,P. A. R. Ade,R. V. Sudiwala
DOI: https://doi.org/10.48550/arXiv.2206.15151
2022-06-30
Instrumentation and Methods for Astrophysics
Abstract:Far-infrared detectors for future cooled space telescopes require ultra-sensitive detectors with optical noise equivalent powers of order 0.2 aW/\sqrt Hz. This performance has already been demonstrated in arrays of transition edge sensors. A critical step is demonstrating a method of fabrication and assembly that maintains the performance but that is extendable to create large-scale arrays suitable, for example, for application in dispersive spectrometers where it may be advantageous to fabricate the array from smaller sub-arrays. Critical here are the methods of assembly and metrology that maintain the required tolerances on the spatial alignment of the components in order to maintain overall performance. These are discussed and demonstrated.