A New Contactless Method For Velocity Measurement Of Bubbleand Slug In Millimeter-Scale Pipelines

Junchao Huang,Haifeng Ji,Zhiyao Huang,Baoliang Wang,Haiqing Li
DOI: https://doi.org/10.1109/ACCESS.2017.2713809
IF: 3.9
2017-01-01
IEEE Access
Abstract:Combining (CD)-D-4 technique and cross-correlation velocity measurement technique, a new measurement method, which is suitable for the bubble/slug velocity measurement in millimeter-scale pipelines, is proposed. Based on the series resonance principle and the simulated inductor technique, a new (CD)-D-4 sensor is developed. With two conductance signals obtained by two new (CD)-D-4 sensors (the upstream sensor and the downstream sensor), the bubble/slug velocity measurement is implemented by the cross-correlation velocity measurement technique. Experiments are carried out in three pipelines with different inner diameters of 4.50, 5.46, and 6.44 mm, respectively. The experimental results show that the proposed bubble/slug velocity measurement method is effective, the development of the new (CD)-D-4 sensor is successful, and the velocity measurement accuracy is satisfactory. The relative error of bubble velocity measurement is less than 5.41% and the relative error of slug FLow velocity measurement is less than 4.90%.
What problem does this paper attempt to address?