Magnetic Barcode Imaging for Contrast Agents.

Andy H Hung,Laura M Lilley,Fengqin Hu,Victoria S R Harrison,Thomas J Meade
DOI: https://doi.org/10.1002/mrm.26175
2016-01-01
Magnetic Resonance in Medicine
Abstract:PURPOSE:To demonstrate a new MR imaging approach that unambiguously identifies and quantitates contrast agents based on intrinsic agent properties such as r1 , r2 , r2*, and magnetic susceptibility. The approach is referred to as magnetic barcode imaging (MBI).METHODS:Targeted and bioresponsive contrast agents were imaged in agarose phantoms to generate T1 , T2 , T2*, and quantitative susceptibility maps. The parameter maps were processed by a machine learning algorithm that is trained to recognize the contrast agents based on these parameters. The output is a quantitative map of contrast agent concentration, identity, and functional state.RESULTS:MBI allowed the quantitative interpretation of intensities, removed confounding backgrounds, enabled contrast agent multiplexing, and unambiguously detected the activation and binding states of bioresponsive and targeted contrast agents.CONCLUSION:MBI has the potential to overcome significant limitations in the interpretation, quantitation, and multiplexing of contrast enhancement by MR imaging probes. Magn Reson Med 77:970-978, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
What problem does this paper attempt to address?