A Machine-Independent Method to Have Active Removal of 5,000 Centistokes Silicone Oil Using Plastic Infusion Tube and 23-Gauge Microcannulas

Zhaotian Zhang,Yantao Wei,Xintong Jiang,Suo Qiu,Shaochong Zhang
DOI: https://doi.org/10.1186/s12886-015-0103-2
IF: 2.086
2015-01-01
BMC Ophthalmology
Abstract:Background To describe one modified method of having machine-independent removal of 5,000 centistokes silicone oil through 23-gauge trocar-cannulas. Methods Consecutive patients with silicone oil tamponade for more than four months and with complete retinal reattachment were included. Two 23-gauge trocars were used to make sclerotomies while the microcannulas remained in situ for intravitreous infusion and silicone oil drainage. A short section of infusion tube was connected with a 10 ml syringe’s needle adapter. The other side was attached to the conjunctiva surface and covered the cannula’s cap inside to form a closed space for silicone oil drainage. The main outcomes were duration for complete removal of silicone oil and intra- and postoperative complications. Result There were totally twenty cases (20 eyes) included. The mean time for draining out the silicone oil was 4.54 ± 0.78 minutes. Intraoperatively, flute needle was introduced additionally in seven cases to achieve complete removal. No cases experienced postoperative visual acuity deterioration or refractory hypotony. No significant residual oil bubbles were observed. No retinal redetachment occurred throughout the follow-ups. Conclusion The modified method of using an infusion tube and 23-gauge trocar-cannulas can achieve quick and complete removal of high viscosity silicone oil.
What problem does this paper attempt to address?