Development Of A Piston-Cylinder Pressure Cell Using Nonmagnetic And High-Yield Strength Alloys For A Squid Magnetometer

ryuji nakai,fumihiko nakamura,takashi suzuki
DOI: https://doi.org/10.1143/JPSJS.76SA.219
2007-01-01
Journal of the Physical Society of Japan
Abstract:A piston-cylinder pressure cell for a commercial SQUID magnetometer (Quantum Design) has been developed by using nonmagnetic and high-yield strength alloys. The cell allows us to achieve pressures of 2.2 GPa at room temperature and 1.7 GPa at low temperatures. We have measured magnetization of paramagnetic palladium and superconducting lead placed within our cell and have estimated that a magnetization of over 2 x 10(-3) emu is possible to measure within the NiCrAl cell as absolute accuracy. As a demonstration of the cell, the paramagnetic susceptibility of pressurised Ca2RuO4 under 1.5 GPa is shown.