AMS Measurement of 32si at the China Institute of Atomic Energy (CIAE)

Ming He,Xianggao Wang,Guowen Zhen,Jie Gong,Xiangdong Ruan,Chaoli Li,Kejun Dong,Zhenyu Li,Hao Hu,Shan Jiang
DOI: https://doi.org/10.1016/j.nimb.2012.01.018
2013-01-01
Abstract:A method for AMS measurement of 32Si has been established at CIAE with the HI-13 Tandem Accelerator and the newly developed ΔE-Q3D detection system. The production of 32Si, the preparation of a set of 32Si reference samples, and their subsequent analysis with the new system are described in detail. An overall suppression factor of more than 1012 for the interfering isobar 32S, resulting in a detection sensitivity below 32Si/Si=5×10−15, has been demonstrated for a limited set of artificially produced samples.
What problem does this paper attempt to address?