Characterization and mechanical investigation of Ti–O 2− x film prepared by plasma immersion ion implantation and deposition for cardiovascular stents surface modification

dong xie,guojiang wan,manfred f maitz,yifeng lei,nan huang,hong sun
DOI: https://doi.org/10.1016/j.nimb.2012.07.033
2012-01-01
Abstract:Up to date, materials for cardiovascular stents are still far from satisfactory because of high risk of biomaterials-associated restenosis and thrombosis. Extensive efforts have been made to improve the biocompatibility of the materials by various surface modification techniques. Ti-O2-x films prepared by plasma immersion ion implantation and deposition (PIII&D) have shown good blood compatibility. For clinical application, surface quality and mechanical durability of the Ti-O2-x film on stents are also of critical importance for the long-term serving. In this paper we present our research results on surface quality, mechanical investigation and characterization of Ti-O2-x films prepared using PIII&D on stent products provided by Boston Scientific SCIMED. Ti-O2-x films with mostly Rutile and little non-stoichiometric phases were obtained with smoothness of <3 nm RMS, largely homogeneity as well as good intergradient film/substrate interface. The Ti-O2-x films on stents products were sustained balloon-expansion of clinically-required extent without mechanical failure, showing highly potential feasibility for cardiovascular stents application. (C) 2012 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?