Construction and Calibration of the Multi-Neutron Correlation Spectrometer at Peking University

H. B. You,Z. H. Yang,Y. L. Ye,Z. H. Li,Y. C. Ge,J. L. Lou,Q. T. Li,J. Xiao,R. Qiao,Z. Y. Tian,Y. L. Sun,H. N. Liu,J. Chen,J. Wu,B. Yang,K. Yang,J. S. Wang,Y. Y. Yang,P. Ma,J. B. Ma,S. L. Jin,J. L. Han
DOI: https://doi.org/10.1016/j.nima.2013.06.037
2013-01-01
Abstract:A Multi-Neutron Correlation Spectrometer (MuNCoS) composed of 80 detector modules was constructed at Peking University. A position resolution of less than 3cm (FWHM) for a typical long scintillation module was obtained based on cosmic ray tests. A calibration method using an unstable nucleus beam of 11Be, together with the fragment identification by using a zero degree telescope, has been successfully developed. Neutron detection efficiencies of 15(1)% and 17(2)% were achieved for neutron energy intervals of 30–40MeV and 40–60MeV, respectively, and at a signal charge threshold of 4MeVee. The experimentally determined neutron detection efficiencies agree well with the simulation based on the GEANT4 code.
What problem does this paper attempt to address?