Glass transition temperature of polymers in supercritical carbon dioxide
Shuming YE,Kai JIANG,Chunyue JIANG,Qinmin PAN
DOI: https://doi.org/10.3321/j.issn:0438-1157.2009.01.039
2009-01-01
Abstract:A novel apparatus for in situ optical measurement of glass transition temperature (T g) was developed. The apparatus consisted of a high-pressure cell equipped with quartz windows, a data acquisition system, a heating and temperature-controlling system, a pressure-controlling system, and a CO 2 feeding system. The volume changes of poly-(ethylene-terephthalate) (PET) when it was swollen by CO 2 were measured in real-time by the apparatus, at the temperatures of 50, 60, 70, 80, 90, 100 and 120°C, and under the pressures of 0.1, 5, 10, 15, 20 MPa of CO 2. Consequently, the glass transition temperatures in SC-CO 2 were obtained under various conditions as indicated above. To examine the accuracy and reliability of this method, comparison of the results obtained by the apparatus under atmospheric pressure with those obtained by the standard differential scanning calorimetry (DSC) were made, and these two methods produced results within acceptable precision (1.6%). Thus the method developed in this work can be reliably used to measure the glass transition temperature of polymers in SC-CO 2.