DETECTORS AND EXPERIMENTAL METHODS: Measurement of the neutron spectrum of a Pu-C source with a liquid scintillator

songlin wang,hanxiong huang,xichao ruan,xia li,jie bao,yangbo nie,qiping zhong,zuying zhou,xiangzhong kong
DOI: https://doi.org/10.1088/1674-1137/33/5/012
IF: 2.944
2009-01-01
Chinese Physics C
Abstract:The neutron response function for a BC501A liquid scintillator (LS) has been measured using a series of monoenergetic neutrons produced by the p-T reaction. The proton energies were chosen such as to produce neutrons in the energy range of 1 to 20 MeV. The principles of the technique of unfolding a neutron energy spectrum by using the measured neutron response function and the measured Pulse Height (PH) spectrum is briefly described. The PH spectrum of neutrons from the Pu-C source, which will be used for the calibration of the reactor antineutrino detectors for the Daya Bay neutrino experiment, was measured and analyzed to get the neutron energy spectrum. Simultaneously the neutron energy spectrum of an Am-Be source was measured and compared with other measurements as a check of the result for the Pu-C source. Finally, an error analysis rind a discussion of the results are given.