Ordered Mesoporous Carbon and Thiolated Polyaniline Modified Electrode for Simultaneous Determination of Cadmium(II) and Lead(II) by Anodic Stripping Voltammetry

lin tang,jun chen,guangming zeng,yuan zhu,yi zhang,yaoyu zhou,xia xie,guide yang,sheng zhang
DOI: https://doi.org/10.1002/elan.201400350
IF: 3.077
2014-01-01
Electroanalysis
Abstract:The fabrication and evaluation of a glassy carbon electrode (GCE) modified with ordered mesoporous carbon (OMC), 2-mercaptoethanesulfonate (MES)tethered polyaniline (PANI) and bismuth for simultaneous determination of trace Cd2+ and Pb2+ by differential pulse anodic stripping voltammetry (DPASV) are presented here. The morphology and electrochemical properties of the fabricated electrode were respectively characterized by scanning electron microscopy (SEM) and electrochemical impedance spectroscopy (EIS). Experimental parameters such as PANI disposition, preconcentration potential, preconcentration time and bismuth concentration were optimized. Under optimum conditions, the fabricated electrode exhibited linear calibration curves ranged from 1 to 120 nM for Cd2+ and Pb2+. The limits of detection (LOD) were 0.26 nM for Cd2+ and 0.16 nM for Pb2+ (S/N=3), respectively. Additionally, repeatability, reproducibility, interference and application were also investigated, and the proposed electrode exhibited excellent performance. The proposed method could be extended for the development of other new sensors for heavy metal determination.
What problem does this paper attempt to address?