Non-enzymatic sensing of uric acid using a carbon nanotube ionic-liquid paste electrode modified with poly(β-cyclodextrin)

Yonghong Li,Xiurong Zhai,Haibo Wang,Xinsheng Liu,Le Guo,Xiaoling Ji,Ling Wang,Hongyan Qiu,Xiaoying Liu
DOI: https://doi.org/10.1007/s00604-015-1522-8
2015-01-01
Mikrochimica Acta
Abstract:We describe a nonenzymatic electrochemical sensor for uric acid. It is based on a carbon nanotube ionic-liquid paste electrode modified with poly(β-cyclodextrin) that was prepared in-situ by electropolymerization. The functionalized multi-walled carbon nanotubes and the surface morphology of the modified electrodes were characterized by transmission electronic microscopy and scanning electron microscopy. The electrochemical response of uric acid was studied by cyclic voltammetry and linear sweep voltammetry. The effects of scan rate, pH value, electropolymerization cycles and accumulation time were also studied. Under optimized experimental conditions and at a working voltage of 500 mV vs. Ag/AgCl (3 M KCl), response to uric acid is linear in the 0.6 to 400 μΜ and in the 0.4 to 1 mΜ concentration ranges, and the detection limit is 0.3 μΜ (at an S/N of 3). The electrode was successfully applied to the detection of uric acid in (spiked) human urine samples. Graphical Abstract SEM images of ( a ) carbon ionic liquid electrode (CILE) ( b ) MWNT-CILE ( c ) β-CD/CILE ( d ) β-CD/ MWNT-CILE. The surfaces of carbon ionic liquid electrode (CILE) ( a ) and MWNT-CILE ( b ) were homogenous and no separated carbon layers can be observed; After β- cyclodextrin (CD) was modified on CILE and MWNT-CILE, the surfaces of β-CD modified electrodes ( c and d ) exhibited loose and porous morphologies.
What problem does this paper attempt to address?