Performance of a microfluidic device for in situ ToF-SIMS analysis of selected organic molecules at aqueous surfaces

li yang,zihua zhu,xiaoying yu,s thevuthasan,james p cowin
DOI: https://doi.org/10.1039/c3ay26513g
IF: 3.1
2013-01-01
Analytical Methods
Abstract:In this study, we report new results concerning the analytical performance of a novel portable vacuum compatible device enabling in situ study of aqueous surfaces using vacuum-based surface analysis tools. The surfaces of aqueous solutions of three representative organic molecules (formic acid, glycerol, and glutamic acid) were analyzed using time-of-flight secondary ion mass spectrometry (ToF-SIMS). Their molecular signals were successfully observed. The device can be operated without interruption in vacuum for up to 8 hours, and SIMS measurements are feasible at any time in this time range. The stability testing of our device under primary ion beam bombardment shows that high fluence (6 x 10(12) ions per cm(2) s(-1)) measurements can be operated continuously for up to 30 minutes without any significant damage to the aperture. However, extra-high fluence measurements (>1 x 10(14) ions per cm(2) s(-1)) may lead to rapid boiling in the aperture, and the aqueous solutions may spread out quickly. Device reproducibility is studied for both consecutive measurements over a short period of time (e.g., 5 min) and intermittent measurements over a long time (e. g., several hours). The relative standard deviation (RSD) for molecular ion signals was determined to be less than 15% for consecutive measurements in 5 min. As to total counts, the RSD is determined to be less than 1% for each chemical compound. Higher RSDs of +/- 40-50% were obtained for intermittent measurements in a few hours, both acceptable for semi-quantitative analysis. In addition, the detection limits of formic acid, glycerol, and glutamic acid are estimated to be 0.04%, 0.008%, and 0.002% (weight ratio), respectively.
What problem does this paper attempt to address?