Acoustic Microstreaming Applied to Batch Micromixing

R Manasseh,K Petkovic-Duran,P Tho,YG Zhu,A Ooi
DOI: https://doi.org/10.1117/12.668030
2005-01-01
Abstract:Experiments are presented in which acoustic microstreaming is investigated and applied to a batch micromixing case appropriate to a point-of-care pathology screening test. The flows presented can be created without complex engineering of contacts or surfaces in the microdevice, which could thus be made disposable. Fundamental flow patterns are measured with a micro-Particle-Image Velocimetry (micro-PIV) system, enabling a quantification of the fluiddynamical processes causing the flows. The design of micromixers based on this principle requires a quantification of the mixing. A simple technique based on digital image processing is presented that enables an assessment of the improvement in mixing due to acoustic microstreaming. The digital image processing technique developed was shown to be non-intrusive, convenient and able to generate useful quantitative data. Preliminary indications are that microstreaming can at least halve the time required to mix quantities of liquid typical of a point-of-care test, and significantly greater improvements seem feasible.
What problem does this paper attempt to address?