Determination of sufentanil in human plasma by LC-MS/MS

曾晓晖,石磊,赵树进,张兴安,谭建坤,刘礼胜
DOI: https://doi.org/10.15887/j.cnki.13-1389/r.2022.19.052
2010-01-01
Abstract:AIM To develop a LC-MS/MS assay to determine sufentanil in human plasma.METHODS Fentanyl was used as an interal standard.Plasma samples were precipitated by acetonitril.The separation was carried out on an Ultimate XB C18(100 mm×2.1 mm,3.0 μm) column at 40℃.The mobile phase consisted of water-methanol solution(15∶85,V/V) with 10 mmol·L-1 ammonium acetate at a flow-rate of 0.2 mL·min-1.Electrospray ionization(ESI) source was applied and operated in the positive ion mode.Quantitaion was performed using multiple reaction monitoring(MRM) of the m/z 387.2→238.1 for sufentanil,and m/z 337.2→188.2 for fentanyl.RESULTS The liner calibration curves were obtained in the range of 0.05-2 μg·L-1 for sufentanil.The extraction recovery for sufentanil was ranged from 83.68%-88.06%.The relative standard deviation of the intraassay and inter-assay precision of variation was less than 10%.CONCLUSION It is an accurate,sensitive and rapid method that can be applied to the evaluation of sufentanil target,and suitable for clinical pharmacokinetic study of sufentanil.