An InGaAs Detector Based Radiation Thermometer and Fixed-Point Blackbodies for Temperature Scale Realization at NIM

X. Hao,Z. Yuan,J. Wang,X. Lu
DOI: https://doi.org/10.1063/1.4819617
2013-01-01
AIP Conference Proceedings
Abstract:In this paper, we describe an InGaAs detector based radiation thermometer (IRT) and new design of fixed-point blackbodies, including Sn, Zn, Al and Cu, for the establishment of a temperature scale from 200 degrees C to 1085 degrees C at the National Institute of Metrology of China. The construction and calibration of the IRT with the four fixed-point blackbodies are described. Characteristics of the IRT, such as the size-of-source effect, the amplifier performance and its stability are determined. The design of the four fixed-points, with 10 mm diameter of aperture and 0.9999 emissivity, is described. The uncertainty of the scale realization is elaborated.