Extended Volume Coverage in Helical Cone-Beam Ct by Using Pi-Line-Based Bpf Algorithm

Seungryong Cho,Xiaochuan Pan
DOI: https://doi.org/10.1117/12.713237
2007-01-01
Abstract:We compared data requirements of filtered-backprojection (FBP) and backprojection-filtration (BPF) algorithms based on PI-lines in helical cone-beam CT. Since the filtration process in FBP algorithm needs all the projection data of PI-lines for each view, the required detector size should be bigger than the size that can cover Tam-Danielsson (T-D) window to avoid data truncation. BPF algorithm, however, requires the projection data only within the T-D window, which means smaller detector size can be used to reconstruct the same image than that in FBP. In other words, a longer helical pitch can be obtained by using BPF algorithm without any truncation artifacts when a fixed detector size is given. The purpose of the work is to demonstrate numerically that extended volume coverage in helical cone-beam CT by using PI-line-based BPF algorithm can be achieved.