Performance Evaluation of the Backprojection Filtered (BPF) Algorithm in Circular Fan-Beam and Cone-Beam CT

LI Liang,CHEN Zhi-Qiang,ZHANG Li,KANG Ke-Jun
DOI: https://doi.org/10.1016/s1001-8042(06)60022-5
2006-01-01
Abstract:In this article we introduce an exact backprojection filtered (BPF) type reconstruction algorithm for cone-beam scans based on Zou and Pan's work. The algorithm can reconstruct images using only the projection data passing through the parallel PI-line segments in reduced scans. Computer simulations and practical experiments are carried out to evaluate this algorithm. The BPF algorithm has a higher computational efficiency than the famous FDK algorithm. The BPF algorithm is evaluated using the practical CT projection data on a 450 keV X-ray CT system with a flat-panel detector (FPD). From the practical experiments, we get the spatial resolution of this CT system. The algorithm could achieve the spatial resolution of 2.4 lp/mm and satisfies the practical applications in industrial CT inspection.
What problem does this paper attempt to address?