Hollow bent output devices in CO2 laser light delivery system

katsumasa iwai,yiwei shi,yuji matsuura,mitsunobu miyagi,junko oishi,yoshihide okagami
DOI: https://doi.org/10.1117/12.463808
2002-01-01
Abstract:In order to change the output direction of laser light in a small operation spare, two types of hollow bent output devices are proposed for medical laser light delivery system. One is the hollow fiber insert-type, which is of the bending radius of 40 mm. The other type, based on a fixed-bent pyrex glass tube, is of the minimum bending radius of 2 mm. Both types of devices are inner-coated with a silver layer and a cyclic olefin polymer (COP) layer. Loss properties for CO2 laser light are clarified for the insert-type devices, called laryngo-tip. And the transmission properties for pilot beams are experimentally discussed. For the fixed-bent tips, the losses are around 0.7 dB, which are almost independent on the bending angles and radii. Transmissions of the tips remain undamaged after 1-hour-delivery of 5 W Output CO2 laser light.
What problem does this paper attempt to address?