Headspace GC-MS Detection of Carbon Monoxide in Decomposed Blood and Hepatic Tissues

Hao H,Zhou H,Zen L,Zhang Z,Yu Z
DOI: https://doi.org/10.15744/2348-9804.1.302
2013-01-01
Abstract:Here in a systematic, accurate and reliable method, Head-Space Gas Chromatography-Mass Spectrometry (HS-GC/MS) was developed to determine blood carboxyhemoglobin (COHb%), in order to investigate deaths related to CO exposure especially involving blood and hepatic tissues.Using a column packed with molecular sieve, COHb levels were quantified down to 0.2% in small blood samples quickly and showed good reproducibility with RSD of the COHb <1%.COHb% in hepatic samples stored at different temperatures (-20 °C for 1-2 years, 0 °C, and 18 °C for two months) can be determined even when the samples are decomposed.The 3-min procedure requires only 0.25 mL of blood sample or 1.0 g of hepatic tissue each time.The technique has a clear advantage over other methods such as UV spectrophotometry.