Digital Pulse Shape Discrimination Methods for N-Γ Separation in an EJ-301 Liquid Scintillation Detector

Wan Bo,Zhang Xue-Ying,Chen Liang,Ge Hong-Lin,Ma Fei,Zhang Hong-Bin,Ju Yong-Qin,Zhang Yan-Bin,Li Yan-Yan,Xu Xiao-Wei
DOI: https://doi.org/10.1088/1674-1137/39/11/116201
2015-01-01
Abstract:A digital pulse shape discrimination system based on a programmable module NI-5772 has been established and tested with an EJ-301 liquid scintillation detector. The module was operated by running programs developed in LabVIEW, with a sampling frequency up to 1.6 GS/s. Standard gamma sources Na-22, Cs-137 and Co-60 were used to calibrate the EJ-301 liquid scintillation detector, and the gamma response function was obtained. Digital algorithms for the charge comparison method and zero-crossing method have been developed. The experimental results show that both digital signal processing (DSP) algorithms can discriminate neutrons from gamma-rays. Moreover, the zero-crossing method shows better n-gamma discrimination at 80 keVee and lower, whereas the charge comparison method gives better results at higher thresholds. In addition, the figure-of-merit (FOM) for detectors of two different dimensions were extracted at 9 energy thresholds, and it was found that the smaller detector presented better n-gamma separation for fission neutrons.