Development of Indigenous Pulse-Shape Discrimination Algorithm for Organic Scintillation detectors

Annesha Karmakar,G. Anil Kumar,Bhavika,V. Anand,Anikesh Pal
2023-07-20
Abstract:The use of programmable hardware devices is imperative for digital based pulse shape discrimination (PSD) to differentiate between various types of radiation. This work reports the development of a PSD algorithm based on tail area and total area, eliminating the need for programmable hardware. The pulses were collected using BC501 detector and Pu-Be source from a digitizer in the oscilloscope mode. The algorithm performs crucial functions such as pulse normalization, shaping, identification and removal of multiple peaks and threshold determination. The algorithm provides neutron and gamma-ray counts, scatter plot, and FoM. In order to test the efficacy of our proposed algorithm, pulses were collected from a different source-detector setup comprising BC501A detector and an Am-Be source from a digitizer in the oscilloscope mode and Charge Integration (CI) mode. The results obtained from our proposed algorithm and CI method clearly indicates a good agreement in terms of number of neutrons and gamma-rays and Figure-of-Merit (FoM), thus providing cost-effective alternative method for neutron and gamma-ray discrimination, offering flexibility and accuracy without specialized hardware.
Instrumentation and Detectors
What problem does this paper attempt to address?