BIOLOGICAL EFFECTS OF THE ARTIFICIAL ANAL SPHINCTER IN THE GOAT MODEL

yasuyuki shiraishi,xiumin zhang,yun luo,hongjian liu,tomoyuki yambe
DOI: https://doi.org/10.1142/9781848169067_0071
2012-01-01
Abstract:Artificial anal sphincters could provide an alternative solution for severe anal incontinence when conventional surgical methods are not possible. We have created an artificial anal sphincter using a shape memory alloy (SMA) and evaluated its validity. The aim of this study was to determine the biological effects and safety after implantation of the SMA artificial anal sphincter in an animal model. The artificial anal sphincter consists of two all-round SMA plates as the main functional parts which perform two basic functions: occlusion at normal body temperature and release upon heating. The SMA artificial anal sphincter was implanted into three female goats under anesthesia. Blood was collected postoperatively from the jugular vein on days 1 and 4, at week 1, and months 1, 2, and 3, and biochemical parameters were examined. Biochemical test results were within normal limits during the 3-month study period. All biochemical measurements of organ function remained within normal range. The results demonstrated that the SMA artificial anal sphincter caused no significant failure of liver, renal, autonomic or digestive functions or damage to smooth muscle. This long-term animal experiment supported the concept and design of implantation with an SMA artificial anal sphincter, and the results demonstrate very good biocompatibility and effective system performance.