Intraocular Lens Power Estimation by Accurate Ray Tracing for Eyes Underwent Previous Refractive Surgeries

Que Yang,Shanshan Wang,Kai Wang,Chunyu Zhang,Lu Zhang,Qingyu Meng,Qiudong Zhu
DOI: https://doi.org/10.1117/12.2193057
2015-01-01
Abstract:For normal eyes without history of any ocular surgery, traditional equations for calculating intraocular lens (IOL) power, such as SRK-T, Holladay, Higis, SRK-II, et al., all were relativley accurate. However, for eyes underwent refractive surgeries, such as LASIK, or eyes diagnosed as keratoconus, these equations may cause significant postoperative refractive error, which may cause poor satisfaction after cataract surgery. Although some methods have been carried out to solve this problem, such as Hagis-L equation([1]), or using preoperative data (data before LASIK) to estimate K value([2]), no precise equations were available for these eyes. Here, we introduced a novel intraocular lens power estimation method by accurate ray tracing with optical design software ZEMAX. Instead of using traditional regression formula, we adopted the exact measured corneal elevation distribution, central corneal thickness, anterior chamber depth, axial length, and estimated effective lens plane as the input parameters. The calculation of intraocular lens power for a patient with keratoconus and another LASIK postoperative patient met very well with their visual capacity after cataract surgery.
What problem does this paper attempt to address?