Label-free biosensing using cascaded double-microring resonators integrated with microfluidic channels

yangqing chen,fang yu,chang yang,jinyan song,longhua tang,mingyu li,jianjun he
DOI: https://doi.org/10.1016/j.optcom.2015.01.028
IF: 2.4
2015-01-01
Optics Communications
Abstract:Fast and accurate quantitative measurement of biologically relevant molecules has been demonstrated for medical diagnostics and drug applications in photonic integrated circuits. Herein, we reported a highly-sensitive optical biosensor based on cascaded double-microring resonators. The sensor was integrated with microfluidic channels and investigated with its label-free detection capability. With a wavelength resolution of 0.47nm, the measured binding capacity of the antibody on the surface exhibits reliable detection limit down to 7.10μg/mL using human immunoglobulin G (hIgG).
What problem does this paper attempt to address?