Research on the Lower Detection Efficiency of the RAD7 for 220rn Than for 222Rn

Y. Tan,D. Xiao,Q. Tang,J. Shan,Q. Zhou,X. Liu
DOI: https://doi.org/10.1088/1748-0221/9/06/t06001
2014-01-01
Journal of Instrumentation
Abstract:The measurement sensitivity of the RAD7 to the Rn-220 concentration in its internal cell is 30% less than the sensitivity to Rn-222. This reduction in Rn-220 sensitivity relative to 222Rn sensitivity has been attributed to the decay of Po-216 before it reaches the collection electrode. Analyzing the airflow patterns in the internal cell of RAD7, we propose that this may also be due to a "short-circuit stream" - highly concentrated Rn-220 injected into the internal cell and discharged from another port quickly. Since in the "short-circuit stream", the Rn-220 concentration of outflow is high, the Rn-220 concentration remaining in the internal cell must be low. Therefore, further improvement to the reliability and accuracy of Rn-220 measurements may be attained by the improvement of the measuring chamber geometric factor and the electric field.