The Impact of Short or Long Transcatheter Occluder Waist Lengths on Postprocedure Complete Atrioventricular Block: A Retrospective Study.

yiyun zhou,Yongwen Qin,Xianxian Zhao,xianzhong lang,Ni Zhu,yongyi bai,Xing Zheng,zhaobiao guo,haobo wu
2015-01-01
Journal of invasive cardiology
Abstract:Background. Complete atrioventricular block (cAVB) is considered the most serious adverse event after transcatheter closure of ventricular septal defect (VSD). Objectives. To evaluate the incidence of postprocedure cAVB and its relationship with different waist lengths of transcatheter occluders. Methods. This is a retrospective study of patients with VSD who had been treated with transcatheter closure at the Shanghai Changhai Hospital from December 2001 to December 2010. Patients were treated with short-waist (n = 234) or long-waist occluders (n = 571). All patients were monitored by electrocardiogram (ECG) within 7 days post procedure. Results. Nine patients (3.8%) treated with short-waist occluders suffered from cAVB and 3 patients (1.3%) required permanent pacemaker implantation, compared with 4 patients (0.7%) and 1 patient (0.2%), respectively, in the long-waist occluder group (P<.001). There was a significantly higher incidence of postprocedure incomplete right bundle branch block (IRBBB) in patients treated with short-waist occluders compared with patients treated with long-waist occluders. There were no significant differences in other procedure-related complications between the two groups. Conclusions. Long-waist VSD occluders are beneficial in the prevention of cAVB and the need for pacemaker implantation after transcatheter closure of VSD.
What problem does this paper attempt to address?