Performance Comparison of 1.5-T Endorectal Coil MRI with 3.0-T Nonendorectal Coil MRI in Patients with Prostate Cancer.

Zarine K Shah,Saba N Elias,Ronney Abaza,Debra L Zynger,Lawrence A DeRenne,Michael V Knopp,Beibei Guo,Ryan Schurr,Steven B Heymsfield,Guang Jia
DOI: https://doi.org/10.1016/j.acra.2014.11.007
IF: 5.482
2015-01-01
Academic Radiology
Abstract:RATIONALE AND OBJECTIVES:To compare prostate morphology, image quality, and diagnostic performance of 1.5-T endorectal coil magnetic resonance (MR) imaging (MRI) and 3.0-T nonendorectal coil MRI in patients with prostate cancer.MATERIALS AND METHODS:MR images obtained of 83 patients with prostate cancer using 1.5-T MRI systems with an endorectal coil were compared to images collected from 83 patients with a 3.0-T MRI system. Prostate diameters were measured, and image quality was evaluated by one American Board of Radiology (ABR)-certified radiologist (reader 1) and one ABR-certified diagnostic medical physicist (reader 2). The likelihood of the presence of peripheral zone cancer in each sextant and local extent was rated and compared to histopathologic findings.RESULTS:Prostate anterior-posterior diameter measured by both readers was significantly shorter with 1.5-T endorectal MRI than with 3.0-T MRI. The overall image quality score difference was significant only for reader 1. Both readers found that the two MRI systems provided a similar diagnostic accuracy in cancer localization, extraprostatic extension, and seminal vesicle involvement.CONCLUSIONS:Nonendorectal coil 3.0-T MRI provides prostate images that are natural in shape and that have comparable image quality to those obtained at 1.5 T with an endorectal coil, but not superior diagnostic performance. These findings suggest an opportunity exists for improving technical aspects of the 3.0-T prostate MRI.
What problem does this paper attempt to address?