Monitoring of toxic compounds in air using a handheld rectilinear ion trap mass spectrometer.
Adam Keil,Heriberto Hernandez-Soto,Robert J Noll,Miriam Fico,Liang Gao,Zheng Ouyang,R Graham Cooks
DOI: https://doi.org/10.1021/ac070906o
IF: 7.4
2008-01-01
Analytical Chemistry
Abstract:A miniature, handheld mass spectrometer, based on the rectilinear ion trap mass analyzer, has been applied to air monitoring for traces of toxic compounds. The instrument is battery-operated, hand-portable, and rugged. We anticipate its use in public safety, industrial hygiene, and environmental monitoring. Gaseous samples of nine toxic industrial compounds, phosgene, ethylene oxide, sulfur dioxide, acrylonitrile, cyanogen chloride, hydrogen cyanide, acrolein, formaldehyde, and ethyl parathion, were tested. A sorption trap inlet was constructed to serve as the interface between atmosphere and the vacuum chamber of the mass spectrometer. After selective collection of analytes on the sorbent bed, the sorbent tube was evacuated and then heated to desorb analyte into the instrument. Sampling, detection, identification, and quantitation of all compounds were readily achieved in times of less than 2 min, with detection limits ranging from 800 parts per trillion to 3 parts per million depending on the analyte. For all but one analyte, detection limits were wen below (3.5-130 times below) permissible exposure limits. A linear dynamic range of 1-2 orders of magnitude was obtained over the concentration ranges studied (sub-ppbv to ppmv) for all analytes.