Direct Mass Spectrometry Analysis of Untreated Samples of Ultralow Amounts Using Extraction Nano-Electrospray.

Yue Ren,Jiangjiang Liu,Linfan Li,Morgan N. McLuckey,Zheng Ouyang
DOI: https://doi.org/10.1039/c3ay41149d
IF: 3.1
2013-01-01
Analytical Methods
Abstract:Direct mass spectrometry analysis of untreated samples of volumes as low as 0.2 µL were achieved using fast extraction and nanoESI (electrospray ionization) in a combined fashion. The analytes in dried samples on paper substrates were extracted by organic solvent in a nanoESI tube and ionized with a high voltage applied for generating a spray. The ionization source produced stable signals for different atmospheric pressure interfaces of triple quadrupole instruments. Analysis time more than 20 minutes were available with 10 µL solvent consumed for the entire analysis process. The performance in qualitative and quantitative analysis was characterized with a wide variety of samples. Limits of detection as low as 0.1 ng/mL (corresponding to an absolute amount of 0.05 pg) were obtained for analysis of atrazine in river water, thiabendazole in orange homogenate, and methamphetamine in blood.
What problem does this paper attempt to address?