Monitoring the reduced scattering coefficient of bone tissues on the trajectory of pedicle screw placement using near-infrared spectroscopy.

Yangyang Liu,Yuyan Wang,Zhiyu Qian,Jinzhe Zhao,Xinzhi Cao,Weitao Li
DOI: https://doi.org/10.1117/1.JBO.19.11.117002
IF: 3.758
2014-01-01
Journal of Biomedical Optics
Abstract:Pedicle screw (PS) fixation has been widely used for spine diseases. Scientists and clinicians employ several approaches to navigate PS during operation. We have demonstrated the feasibility of monitoring the reduced scattering coefficient (mu(s)') on the trajectory of PS using near-infrared spectroscopy (NIRS). To perform the in-vitro monitoring, an NIRS measurement system was introduced and the reduced scattering coefficients of different sites in porcine pedicle were accurately deduced from the spectrum. Moreover, the changes of the reduced scattering coefficient along the different paths were studied. The results show reduced scattering coefficients on different regions of bones can be significantly distinguished. Furthermore, monitoring experiments along different paths confirmed that a reduced scattering coefficient would change versus the depth of puncture in pedicles. Thus, the proposed monitoring system based on NIRS provides a potential for guiding PS during operation. (C) The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License.
What problem does this paper attempt to address?