Hemocompatibility research on the micro-structure surface of a bionic heart valve.

Xia Ye,Ze Wang,Xianghua Zhang,Ming Zhou,Lan Cai
DOI: https://doi.org/10.3233/BME-141049
2014-01-01
Bio-Medical Materials and Engineering
Abstract:In order to study how the geometric parameters and shape of the micro-structure surface of a bionic heart valve affects hemocompatibility, mastoid micro-structures with different periodic space were fabricated using a femtosecond laser on a polyurethane (PU) surface. The apparent contact angles of droplets on the micro-structure surfaces were measured to characterize their wettability. Then a series of blood compatibility experiments, including platelet adhesion, dynamic coagulation and hemolysis were completed. The experimental results showed that the micr-ostructure on the biomaterial surface helped improve its hydrophobicity and hemocompatibility. Also, the periodic space affected not only the hydrophobicity but also the hemocompatibility of the biomaterial. With the increasing of the periodic space, the apparent contact angle increased, the number of platelet adhesion decreased, the dynamic clotting time became longer and the hemolysis ratio reduced. In addition, the shape of the micro-structure also affected the hemocompatibility of the biomaterial.
What problem does this paper attempt to address?