Cleanup with Two Florisil Columns for Gas Chromatographic Determination of Multiple Pyrethroid Insecticides in Products of Animal Origin.

GF PANG,TS ZHAO,YZ CHAO,CL FAN
DOI: https://doi.org/10.1093/jaoac/77.6.1634
1994-01-01
Journal of AOAC International
Abstract:A gas chromatographic method was developed for the simultaneous determination of 9 pyrethroid insecticides in products of animal origin. The multiresidues of the pyrethroids in different samples were extracted with acetone-petroleum ether (1 + 1), and extracts were cleaned up on a Florisil partition column and a Florisil adsorption column. Four animal tissue samples were examined at 0.05-0.25 ppm fortification levels. The average recoveries of all insecticides were 76.9-88.0%, and the coefficients of variation were < 4.6% for all insecticides except permethrin. The detection limits of the method were ca 5 ppb for all insecticides but permethrin, which had a detection limit of ca 10 ppb.