Ion Source And Injection Line For High Intensity Medical Cyclotron

XianLu Jia,Fengping Guan,Hongjuan Yao,TianJue Zhang,Jianjun Yang,Guofang Song,Tao Ge,Jiuchang Qin
DOI: https://doi.org/10.1063/1.4829739
IF: 1.6
2014-01-01
Review of Scientific Instruments
Abstract:A 14 MeV high intensity compact cyclotron, CYCIAE-14, was built at China Institute of Atomic Energy (CIAE). An injection system based on the external H- ion source was used on CYCIAE-14 so as to provide high intensity beam, while most positron emission tomography cyclotrons adopt internal ion source. A beam intensity of 100 mu A/14 MeV was extracted from the cyclotron with a small multi-cusp H- ion source (CIAE-CH-I type) and a short injection line, which the H- ion source of 3 mA/25 keV H- beam with emittance of 0.3 pi mm mrad and the injection line of with only 1.2 m from the extraction of ion source to the medial plane of the cyclotron. To increase the extracted beam intensity of the cyclotron, a new ion source (CIAE-CH-II type) of 9.1 mA was used, with maximum of 500 mu A was achieved from the cyclotron. The design and test results of the ion source and injection line optimized for high intensity acceleration will be given in this paper. (C) 2013 AIP Publishing LLC.
What problem does this paper attempt to address?