Integrated Sample Cleanup-Capillary Electrophoresis Microchip for High-Performance Short Tandem Repeat Genetic Analysis

Stephanie H. I. Yeung,Peng Liu,Nadia Del Bueno,Susan A. Greenspoon,Richard A. Mathies
DOI: https://doi.org/10.1021/ac8018685
IF: 7.4
2008-01-01
Analytical Chemistry
Abstract:An integrated PCR sample cleanup and preconcentration process is developed for forensic short tandem repeat (STR) analysis using a streptavidin-modified photopolymerized capture gel injector for microchip capillary electrophoresis (mu CE). PCR samples generated with one biotinylated primer and one fluorescent primer provide the input to the streptavidin-based affinity capture-mu CE device. Monoplex PCR samples processed by the device exhibited similar to 10- to 50-fold increased fluorescence intensities, and DNA profiles generated using 9-plex STR samples displayed similar to 14- to 19-fold higher signal intensities compared to those analyzed using traditional cross injection. Complete STR profiles were obtained with as few as 25 copies of DNA template using the capture-mu CE device. Four DNA samples with various degrees of degradation were also tested. Samples analyzed using the capture-mu CE device resulted in a significant increase of successful allele detection. The ability of our capture-mu CE device and method to remove contaminating ions, to concentrate the sample injection plug, and to eliminate electrokinetic injection bias provides a powerful approach for integrating sample cleanup with DNA separation.
What problem does this paper attempt to address?