Microfluidic system for high-throughput immunoglobulin-E analysis from clinical serum samples
Lulu Zheng,Yongfeng Fu,Xiran Jiang,Suqin Man,Wei Ran,Meng Feng,Sixiu Liu,Xunjia Cheng,Guodong Sui
DOI: https://doi.org/10.1016/j.talanta.2015.05.014
IF: 6.1
2015-01-01
Talanta
Abstract:Rapid and high-throughput analytical techniques for IgE that requires a small serum amount are very important, especially for pediatric patients. In these patients, blood is collected from veins, which is painful compared to fingertip blood collection. Herein, a novel microfluidic system capable of high-throughput parallel analyses of allergen-specific IgE from small amounts of patient serum was successfully developed. A six-plex immunoassay was constructed within a microfluidic chip, and the entire system was validated using samples from clinical patients. Major antigens from house dust mite (Dermatophagoides farinae and Blomia tropicalis), cat (Felis domesticus), fungus (Cladosporium herbarum), ragweed (Humulus japonicas), and tree pollen (Platanus acerifolia) were used as analysis targets. Sample consumption decreased to <0.05µL compared with the 480µL serum consumption by fluoroenzyme immunoassay (UniCAP system Pharmacia Diagnostics AB, Uppsala, Sweden), the 50µL serum consumption by enzyme-linked immune sorbent assay (ELISA), or the 1.5µL serum consumption by conventional protein chip analysis. Analysis duration, reagent cost, and total cost for each measurement were also considerably decreased. The assay showed good accuracy and sensitivity toward the clinical samples. A significant correlation of allergen-specific IgE levels was found among the microfluidic assay, UniCAP system, and ELISA.